      PERANCANGAN ALAT STERILISATOR MENGGUNAKAN ULTRAVIOLET DAN PANAS KERING

      Thumbnail
      View/Open
      COVER (228.4Kb)
      HALAMAN JUDUL (145.9Kb)
      HALAMAN PERNYATAAN (81.41Kb)
      HALAMAN PERSETUJUAN (88.66Kb)
      KATA PENGANTAR (157.2Kb)
      MOTTO (82.36Kb)
      HALAMAN PERSEMBAHAN (186.2Kb)
      DAFTAR ISI (98.05Kb)
      DAFTAR GAMBAR (87.47Kb)
      DAFTAR TABEL (84.82Kb)
      ABSTRAK (161.0Kb)
      BAB I (159.8Kb)
      BAB II (773.7Kb)
      BAB III (989.5Kb)
      BAB IV (629.8Kb)
      BAB V (139.2Kb)
      DAFTAR PUSTAKA (162.3Kb)
      LAMPIRAN (744.6Kb)
      ANALISA PERHITUNGAN LAPORAN (259.9Kb)
      Date
      2016-11
      Author
      TANTONI, TANTONI
      Metadata
      Show full item record
      Abstract
      Dental dental equipment that is often used to be contaminated with bacteria and microorganisms if the tool does not sterilize. Based on the problems that needed sterilizer, whereas the existing sterilizer is still manual, the temperature used is not in accordance with the sterilization of dental unit and non-metallic equipment can not be sterilized with high temperatures. Writers interested in creating a sterilizer utilizing a heating element with a temperature of 1600C suitable for sterilization of dental unit and utilizes ultraviolet light to sterilize dental equipment that is not strong against high temperature, thermocouple as a temperature sensor and ATMega 16 as a controller module. Based on the results of laboratory tests that have been done with the author concluded that, after the unit of dental equipment is sterilized using ultraviolet sterilization contained three bacterial colonies, and dry heat sterilization are two colonies of bacteria. It can be concluded that the author made effective module and is able to kill bacteria in dental unit.
      URI
      http://repository.umy.ac.id/handle/123456789/6466
      Collections
      • Department of Electrical Engineering

      DSpace software copyright © 2002-2015  DuraSpace
      Contact Us | Send Feedback
      Theme by 
      @mire NV
       

       

      Browse

      All of UMY RepositoryCollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

      My Account

      Login

      DSpace software copyright © 2002-2015  DuraSpace
      Contact Us | Send Feedback
      Theme by 
      @mire NV