STIMULATOR SpO2 BERBASIS MIKROKONTROLER ATmega 328
Abstract
Pulse Oximeter or Oxygen Peripheral Capillary Saturation (SPO2) used to monitor the
saturation or oxygen content in the patient's blood (artery) and heart rate in one minute without
surgery or non-invasive. Given the importance of the accuracy and precision of the SPO2 pulse
oximeter in the diagnosis, in the process of treatment and inspection of the equipment, the function
test and calibration of SpO2. SpO2 function and calibration test requires signal input from the
heart and hemoglobin color change. The pulse simulation of the heart and the color change of
hemoglobin can be obtained electronic by utilizing microcontroller, LED, and infra LED to
become a stimulator SpO2. The SpO2 stimulators are cheap and simple with internally-equipped
calibrations. The test result gets an average deviation value of 0.10 BPM with a 0% deviation
percentage, standard deviation indicates an average value of 0.70BPM and uncertainty is
obtained by an average value of 0.02 BPM makes this SpO2 stimulator device very feasible to
use