dc.contributor.advisor | ATMAJA,BAMBANG GIRI | |
dc.contributor.author | PUSPITASARI, INNES DYAH IKA | |
dc.contributor.author | PUSPITASARI, INNES DYAH IKA | |
dc.contributor.author | PUSPITASARI, INNES DYAH IKA | |
dc.date.accessioned | 2016-12-13T05:46:21Z | |
dc.date.available | 2016-12-13T05:46:21Z | |
dc.date.issued | 2016-08-25 | |
dc.identifier.uri | http://repository.umy.ac.id/handle/123456789/7150 | |
dc.description | Kalibrator Suction Pump adalah alat yang tergolong kedalam alat kesehatan yang berfungsi untuk mengkalibrasi alat suction pump pada rumah sakit. Alat ini sangatlah penting untuk rumah sakit karena alat ini dapat mengetahui ketidaklayakan dari alat suction pump itu sendiri. Pengecekan kelayakan alat kesehatan dipayakan untuk lebih teliti dan akurat, maka dari itu diperlukan sebuah alat untuk kalibrasi suction pump dengan keakuratan pengukuran yang sudah melalui tes kelayakan alat dengan menggunakan pressure meter terstandar.
Disini penulis membuat kalibrator suction pump yang hasilnya akan diukur dan dibandingkan dengan Digital Pressure Meter (DPM) yang sudah terstandar, modul menggunakan sistem mikrokontroler dan didisplaykan dengan lcd 16x2.
“Portable Kalibrator Suction Pump Berbasis Mikrokontroler ATMega 16” dibandingkan dengan digital pressure meter untuk mengetahui keakurasian waktu dioperasikan. Nilai rata-rata error yang didapat dari pengkuran pada enam titik (-100mmHg, -200mmHg, -300mmHg, -400mmHg, -500mmHg, -600mmHg) dan 20kali pengukuran dengan yakni sebesar 0,00343677%. | en_US |
dc.description.abstract | serves to calibrate suction pump at the hospital. This tool is very important to
the hospital because this tool can know suction pump inadequacies of the tool
itself. Checking eligibility dipayakan medical devices for more thorough and
accurate, and therefore needed a tool for calibration suction pump with the
accuracy of measurements that have been through the test the feasibility of the
tool by using a pressure meter standardized.
Here the authors make calibrator suction pump which results will be
measured and compared with the Digital Pressure Meter (DPM), which has been
standardized, Modul use the system microcontroller and display with 16x2 lcd.
"Calibrator Portable Suction Pump-Based Microcontroller ATMega 16"
compared to the digital pressure meter to determine the accuracy of the time
operated. The average value obtained from the error of taking the measurements
at six points (-100mmHg, -200mmHg, -300mmHg, -400mmHg, -500mmHg, -
600mmHg) and 20 measurement with which is equal to 0,00343677%.
__________________________________________________________________ | en_US |
dc.publisher | VOC UMY | en_US |
dc.subject | calibrators, Suction Pump, DPM. Kalibrator, Suction Pump, DPM | en_US |
dc.title | PORTABLE KALIBRATOR SUCTION PUMP BERBASIS MIKROKONTROLER ATMEGA 16 | en_US |
dc.type | Image, 3-D
VOC
54 | en_US |