PORTABLE KALIBRATOR SUCTION PUMP BERBASIS MIKROKONTROLER ATMEGA 16
Date
2016-08-25Author
PUSPITASARI, INNES DYAH IKA
PUSPITASARI, INNES DYAH IKA
PUSPITASARI, INNES DYAH IKA
Metadata
Show full item recordAbstract
serves to calibrate suction pump at the hospital. This tool is very important to
the hospital because this tool can know suction pump inadequacies of the tool
itself. Checking eligibility dipayakan medical devices for more thorough and
accurate, and therefore needed a tool for calibration suction pump with the
accuracy of measurements that have been through the test the feasibility of the
tool by using a pressure meter standardized.
Here the authors make calibrator suction pump which results will be
measured and compared with the Digital Pressure Meter (DPM), which has been
standardized, Modul use the system microcontroller and display with 16x2 lcd.
"Calibrator Portable Suction Pump-Based Microcontroller ATMega 16"
compared to the digital pressure meter to determine the accuracy of the time
operated. The average value obtained from the error of taking the measurements
at six points (-100mmHg, -200mmHg, -300mmHg, -400mmHg, -500mmHg, -
600mmHg) and 20 measurement with which is equal to 0,00343677%.
__________________________________________________________________